Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Yokogawa uR1000: CAL ADJ DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 93 ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Connect UUT to AC power and turn unit ON. 1.004 DISP 1.004 DISP Warm-up time: 30min - REQUIRED 1.004 DISP Temperature: 23 +/-2 degC 1.004 DISP Humidity: 55 +/-10% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD ZERO PEN 1.007 DISP Enter the SET UP mode by turning the AC POWER OFF, 1.007 DISP press and HOLD the key, turn the AC POWER ON 1.007 DISP hold key until the SET UP display appears. 1.008 DISP Use the UP/DOWN keys to select the display 1.008 DISP 'SETUP=P ADJ', and press the key. 1.009 DISP Use the UP/DOWN keys to select the display 1.009 DISP 'P ADJ=ZERO', and press the key. 1.009 DISP 1.009 DISP Select the channel #(=pen number) which you want to 1.009 DISP adjust, then press the key. 1.010 DISP Adjust value so the recording position is 0 +/-.2mm. 1.010 DISP 1.010 DISP Changing the numeric value by 1 will result in a pen 1.010 DISP movement of .02mm. The pen moves AFTER pressing . 1.010 DISP 1.010 DISP Repeat procedure until pen is at 0 on the chart grid. 1.011 DISP The display ZERO SET will appear. 1.011 DISP 1.011 DISP Then press the key. 1.011 DISP 1.011 DISP Repeat procedure for any other channel that needs adj. 1.012 OPBR Do you wish instructions for adjusting another channel? 1.013 JMPT 1.008 1.014 HEAD FULL SPAN PEN 1.015 DISP Use the UP/DOWN keys to select the display 1.015 DISP 'SETUP=P ADJ', and press the key. 1.015 DISP Use the UP/DOWN keys to select the display 1.015 DISP 'P ADJ=FULL', and press the key. 1.015 DISP 1.015 DISP Select the channel #(=pen number) which you want to 1.015 DISP adjust, then press the key. 1.016 DISP Adjust value so the recording position is 100 +/-.2mm. 1.016 DISP 1.016 DISP Changing the numeric value by 1 will result in a pen 1.016 DISP movement of .02mm. The pen moves AFTER pressing . 1.016 DISP 1.016 DISP Repeat procedure until pen is at 100 on the chart grid. 1.017 DISP The display FULL SET will appear. 1.017 DISP 1.017 DISP Then press the key. 1.017 DISP 1.017 DISP Repeat procedure for any other channel that needs adj. 1.018 OPBR Would you like instructions again for another channel? 1.019 JMPT 1.015 1.020 DISP Select the 'SETUP=END' display using the UP/DOWN 1.020 DISP keys, and press the key. 1.020 DISP 1.020 DISP Select the 'END=STORE' display and press key. 1.021 DISP After several seconds, the unit will store the 1.021 DISP settings and return to normal operation. 1.022 END